Cytotoxic Safety Cabinet

Technical parameters:

Model

11234BBC86

Internal Size (W*D*H)

1220*600*630mm 48" *23.6"*24.8"

External Size (W*D*H)

1370 *760 *2100 mm 53.9"*29.9" *86.6"

Tested Opening

Safety Height ≤ 200mm

Max Opening

400mm

Internal Work Area, space

0.73 m2

Inflow Velocity

0.46m/s

Down Flow Velocity

0.33m/s

Exhaust Volume with Thimble Duct

CBV Exhaust Volume 611 m3 / h (360 cfm)

Static Pressure at CBV

Exhaust Volume 39 Pa / 0.15 in H2O

Air Supply Filter

ULPA filter, efficiency > 99.999% for particle size between 0.1 -0.2 um

2 pieces ULPA filter, efficiency > 99.999% for particle size between 0.1 -0.2 um

Exhaust Filter

Filter life indicator and Filter working time indicator.

Alarm

Audio and visual alarm, filter replacement, Window over height, Abnormal airflow velocity

Noise

< 60dB

Water & Gas Tap

Water Tap *1, Gas tap*1

UV Lamp

30W*1;

Fluorescent Lamp

28W*2

Emission of 253.7 nanometers for most efficient decontamination,

UV life indicator

Illumination

>1000Lux

Consumption

<700W

Waterproof Socket

Two, Total load of two sockets: <500W

Display

LCD display ,

Control System

Microprocessor

Humidity Temperature

Humidity and Temperature can be displayed on the control panel

Material

work area: 304 stainless steel

Main body: cold-rolled steel sheets with anti-bacteria powder

Front Window

motorized, Two-layer laminated toughened glass≥5mm

caster

Universal caster with brake and leveling feet

Power supply

110/220V±10%, 60/50Hz

Standard Accessory

Base stand, Drain valve, waterproof sockets*2, Remote control, Foot switch, UV lamp*2 and Fluorescent Lamp, Water tap, Gas tap

Optional Accessory

Armrest, Formalin fumigation sterilizer, infrared sterilizer, airflow tester

Gross Weight

350kg

Package size

1570*950*2220mm

 

Download Literature:
Cytotoxic Safety Cabinet

Send Quick Inquiry
Your Name: Required
Your E-mail: Required
Subject: Do Not Change It
Message:
 






























LCD Display
Large display is easy to monitor all the safety parameters at a glance,and ergonomically sized control panel improves users interface.


Foot Switch
Adjust front window height by foot during experiment, to avoid airflow turbulence caused by arm movement.


Water and Gas Tap
High quality stainless steel.


UV Lamp
Emission of 253.7 nanometers for most efficient decontamination.


Front window
Two-layer laminated toughened glass >5mm, motorized.

TEL: +86-25-84677741   FAX: +86-25-84677742  http://www.pioway.com  Mail: info@pioway.com  © 2001-2030  PIOWAY All Rights Reserved.