Hemaray86 5 Diff. Hematology Analyzer with Auto Sampler

Features:
28 parameters, 2 histograms, 2 scattergrams60 samples/hour
Laser scatter technology
Integral titanium incubation system
3-D topographic technology
Constant linear sheath flow
With Auto Sampler

Specifications:

Regular parameters:
WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, PDW, MPV, PCT, P-LCR, NEU#, NEU%, LYM#, LYM%, MON#, EOS#, EOS%, BAS#, BAS%

Research parameters:
ALY#, ALY%, IG#, IG% 2 histograms, 2 scattergrams

Test speed
Close tube mode: 60 samples/hour

Test mode
CBC+5DIFF, CBC

Performanace:

Item

Linear range

Carry over

CV

WBC

1.0-99.9x10 9 /L

< 0.5%

< 2.0%

RBC

0.3-7.0x10 12 /L

< 0.5%

< 1.5%

HGB

20-240g/L

< 0.5%

< 1.5%

PLT

20-999x10 9 /L

< 1.0%

< 4.0%