Features:
1. Pneumatically driven electrically controlled ventilator, can be used for emergency clinical resuscitation and respiratory therapy
2. Display: High visibility, wide angle LED alphanumeric data display, equipped with tactile analog parameter controls
3. Equipped with high precession infrared turbine pressure and flow sensors
4. Power failure support functions: Automatically converts to standby mode during AC power outages
5. Nonhazardous environmentally friendly frame, precise construction, elegant design, equipped with automated self-checking function during startup and an automatic self-calibrating sensor
6. Alarms: Audible and/or visual alarms for tidal volume, airway pressure, power failure oxygen failure and other alarms.
7. 9 Level adjustable humidifier
8. Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Specifications:
Model |
PA500 |
Display |
LED |
Ventilation Mode |
A/C, IPPV, SIPPV, IMV, SIMV, MANUAL |
Minute Volume |
Greater than or equal to 18L |
Back-up Power Supply |
At least 4 hours |
Tidal Volume |
Adjustable range: 50-1500ml, Display range: 0-2000ml |
Output Oxygen Concentration |
<50% |
Respiratory Rate |
6-60bpm |
Pressure Limit |
1-6kPa |
Inspiratory/Expiratory (I:E) Ratio |
1:1.5-1:3.0 |
PEEP |
1-10cmH 2 O |
SIMV Rate |
1-12bpm |
Inspiratory Pressure Trigger |
. -4-10cmH 2 O |
Maximum Safe Operating Pressure |
≤ 6.0 kPa |
SIGH |
1 sigh breath in every 80-120 breaths |
Airway Pressure Alarm |
Upper Limit: 0.9~5.4kPa, Lower Limit: 0.5kPa |
Gas Supply Requirement |
280~600 kPa medical grade oxygen |
|
|