Features:
-UV wavelength available, including 340nm
-End Point, Fixed time, Kinetic
-Temperature control(optional), incubation time and temperature adjustable
-PC controlled system, Windows XP
-8-channel optical fiber system enables 5 secs’ reading for the whole plate
-More than 500 programs and minimum 100,000 results can be stored
-Powerful QC programs including Westguard Multi-rule and Instant method, automatic alarm
-Calculation modes include ABS, Cut-off, Curve, Linear, Log, Exponential, Power and 4PL regression
-Auto self-check when power on, with lamp saving and plate shaking
-Random positive and negative control setting
-Multi-assay enables up to 12 different assays on one plate
Technical Specifications:
General Descriptions
Absorbance Range
0.000-4.000 Abs
Resolution
0.01 Abs(displayed), 0.0001Abs(calculated)
Light Source
Halogen Lamp
Wavelength
340, 405, 450, 492, 630nm, 3 more filters optional
Wavelength Accuracy
±1nm
Band Width
<8nm
Methodologies
End Point, Fixed time, Kinetic
Calculation Method
ABS, Cut-Off, Single Standard, Curve, Multi-Percent, Percent Log, Linear, Exponential, Power, 4PL regression
Reading Speed
5 seconds for 96 well plate(single wavelength)
Plate Incubation
Ambient temperature +4~45°C, adjustable
Interface
RS-232
Operation
PC operation, bidirectional communications
|
|