XI-921 Electrolyte Analyzer

Features
Compact, cost-efficient and easy - to- use
Excellent precision and reliability
Long life, high performance, maintenance free electrodes
Automatic sampling, washing and calibration
Touch screen ,large LCD display
Internal thermal recorder
Auto sampler optional

Specifications
Principle: Direct measurement by Ion Selective Electrode (ISE)
Sample: Whole blood, serum, plasma, diluted urine
Sample volume 150µL
Throughput: up to 60 samples/h
Data storage: up to 200 patient result
Calibration: Automatic or On-demand
Input: Touch-screen
Display: Large LCD with backlight
Output: Internal thermal recorder, RS-232 serial port
Working Environment:Temperature: 15~32 degree
Relative humidity: 85%
Input voltage: AC 220V/110V ±10%, 50/60Hz
Power consumption: 60W
Dimensions: 440mm×380mm×350mm (H×W×D)
Weight Main unit 10kg
Auto sampler (option) 1.5kg

Measuring range

Reproducibility(CV%)

K+

0.50~15.00 mmol/L

<1.5%

Na+

30.0~200.0 mmol/L

<1.0%

Cl-

30.0~200.0 mmol/L

<1.5%

Ca2+

0.10~5.00 mmol/L

<2.0%

Li+

0.20~3.00 mmol/L

<3.0%

pH

4.00~9.00 unit

<1.0%

Simple and reliable design of fluid tubing system and electrodes, minimize the maintenance work

Automatic sampling and probe washing,
Minimize the carry-over
Touch screen , user friendly software ,
easy to operate

Different Models

XI-921F K+ / Na+ / Cl-
XI-921A K+ / Na+
XI-921B K+ / Na+ / Cl- / HCO3- / AG
XI-921C K+ / Na+ / Cl- / Ca2+ / pH
XI-921D K+ / Na+ / Cl- / Ca2+ / pH / HCO3- /AG
XI-921E K+ / Na+ / Cl- / Li+



Download Literature:
XI-921 Series
Right click and save as... to download the literature