Applications
General Photography, Ray-filter Photography
Features
1.Single bed and single bulb.
2.Design according to the electric mechanism of middle-size machine. The function is full and reasonable.
3.The high-tension system is of march-past structure. Equipped with a full wave rectification high-voltage generator and independent X-ray tube unit, with large output power and reliable and stable operation.
4.Supply voltage V, roentgenoscopy, photography KV and stepless adjustable.
5.Equipped with the manostat for filament of X-ray tube and the space charge compensator as to ensure the stable and accurate output.
6.Equipped with the capacity limitation and protection circuit and KvmA sinter lock limitation and protection.
7.Adopt the digital circuit timer, grade according to R10 priority coefficient and can control time exactly
8.Examining bed is of electric back rotation and is equipped with vibrating filter-ray device.
9.The photographic flat bed has a vibrating ray-filter.
10.The upright post has the bottom track only, which is easy while installation.
Main specification:
Power supply |
Adjustable range |
380V: 340V--420V |
220V: 195V--250V |
Radiography |
Adjustable range of tube voltage |
50~100kV (Can adjust continuously) |
Adjustable range of tube current |
Small focus: 50mA
Big focus: 50, 100, 150 and 200mA |
Tube current of spot-film |
150mA (fix) |
Adjustable range of time |
0.04~6.3s 23 grades together |
High-voltage Generator unit |
Max. DC output voltage |
100kV |
Max. DC output current |
200mA at instant 5mA at running |
X-ray tube |
Model |
XD51-20.40/100 |
Focus |
1mm x1mm 2mmx2mm |
Rotary angle |
+90~0~-5 degree |
Move range of gastrointestinal spot-film device (fluorescent screen) |
Pressure direction: >300mm |
Horizontal direction: the center of diagnostic bed moves left (right) >80mm |
Vertical direction: > 400mm |
|
|